Report on method for improved, gravimetric Winkler titration,

View/ Open
Average rating
votes
Date
2013Author
Helm, Irja
Jalukse, Lauri
Leito, Ivo
Status
Published
Metadata
Show full item recordAbstract
This report presents a high-accuracy Winkler titration method that has been developed
for determination of dissolved oxygen concentration. Careful analysis of uncertainty
sources relevant to the Winkler method was carried out and the method was optimized
for minimizing all uncertainty sources as far as practical. The most important
improvements were: gravimetric measurement of all solutions, pre-titration to
minimize the effect of iodine volatilization, accurate amperometric end point detection
and careful accounting for dissolved oxygen in the reagents. As a result, the developed
method is possibly the most accurate method of determination of dissolved oxygen
available. Depending on measurement conditions and on the dissolved oxygen
concentration the combined standard uncertainties of the method are in the range of
0.012 – 0.018 mg dm-3 corresponding to the k = 2 expanded uncertainty in the range of
0.023 – 0.035 mg dm-3 (0.27 – 0.38%, relative). This development enables mo.....
Publisher
University of Tartu, Institute of Chemistry,Tartu, Estonia
Document Language
enSustainable Development Goals (SDG)
14.aEssential Ocean Variables (EOV)
OxygenSpatial Coverage
EuropeanCitation
Helm, I., Jalukse, L. and Leito I. (2013) Report on method for improved, gravimetric Winkler titration. Tartu, Estonia University of Tartu, Institute of Chemistry, 38pp. DOI: http://dx.doi.org/10.25607/OBP-1231Collections