Show simple item record

dc.contributor.authorHelm, Irja
dc.contributor.authorJalukse, Lauri
dc.contributor.authorLeito, Ivo
dc.coverage.spatialEuropean
dc.date.accessioned2021-06-29T18:31:33Z
dc.date.available2021-06-29T18:31:33Z
dc.date.issued2013
dc.identifier.citationHelm, I., Jalukse, L. and Leito I. (2013) Report on method for improved, gravimetric Winkler titration. Tartu, Estonia University of Tartu, Institute of Chemistry, 38pp. DOI: http://dx.doi.org/10.25607/OBP-1231en_US
dc.identifier.urihttps://repository.oceanbestpractices.org/handle/11329/1599
dc.identifier.urihttp://dx.doi.org/10.25607/OBP-1231
dc.description.abstractThis report presents a high-accuracy Winkler titration method that has been developed for determination of dissolved oxygen concentration. Careful analysis of uncertainty sources relevant to the Winkler method was carried out and the method was optimized for minimizing all uncertainty sources as far as practical. The most important improvements were: gravimetric measurement of all solutions, pre-titration to minimize the effect of iodine volatilization, accurate amperometric end point detection and careful accounting for dissolved oxygen in the reagents. As a result, the developed method is possibly the most accurate method of determination of dissolved oxygen available. Depending on measurement conditions and on the dissolved oxygen concentration the combined standard uncertainties of the method are in the range of 0.012 – 0.018 mg dm-3 corresponding to the k = 2 expanded uncertainty in the range of 0.023 – 0.035 mg dm-3 (0.27 – 0.38%, relative). This development enables more accurate calibration of electrochemical and optical dissolved oxygen sensors for routine analysis than has been possible before. Most of this report is based on the article I.Helm, L.Jalukse, I. Leito, Anal. Chim. Acta. 741 (2012) 21-31.en_US
dc.description.sponsorshipEuropean Metrology Research Programme (EMRP)en_US
dc.language.isoenen_US
dc.publisherUniversity of Tartu, Institute of Chemistry,en_US
dc.subject.otherWinkler methoden_US
dc.subject.otherDissolved oxygenen_US
dc.subject.otherUncertainty quantificationen_US
dc.subject.otherPrimary methoden_US
dc.titleReport on method for improved, gravimetric Winkler titration,en_US
dc.typeReporten_US
dc.description.statusPublisheden_US
dc.publisher.placeTartu, Estoniaen_US
dc.subject.parameterDisciplineDissolved gasesen_US
dc.description.currentstatusCurrenten_US
dc.description.sdg14.aen_US
dc.description.eovOxygenen_US
dc.description.adoptionValidated (tested by third parties)en_US
dc.description.methodologyTypeMethoden_US
dc.description.methodologyTypeDescription of a metrology standarden_US
obps.endorsementAuthorDeclared.recommendedPracticeEuropean Metrology Research Programme (EMRP)
obps.contact.contactnameIvo Leito
obps.contact.contactemailivo.leito@ut.ee
obps.resourceurl.publisherhttps://sisu.ut.ee/sites/default/files/measurement/files/g_winkler_report_280613_0.pdf