Performance Verification Statement for the Chelsea UviLux Hydrocarbon and CDOM Fluorometers.

View/ Open
Average rating
votes
Date
2013Author
Johengen, T.
Smith, G.J.
Purcell, H.
Loranger, S.
Gilbert, S.
Maurer, T.
Gundersen, K.
Robertson, C.
Tamburri, M.
Status
PublishedPages
58pp.
Metadata
Show full item recordAbstract
Instrument performance verification is necessary so that effective existing technologies can
be recognized, and so that promising new technologies can become available to support coastal
science, resource management, and ocean observing systems. The Alliance for Coastal Technologies
(ACT) has therefore completed an evaluation of commercially available in situ hydrocarbon sensors.
This verification included test applications for: (1) controlled laboratory tanks with additions of
various organic, fluorescent compounds, (2) experimental wave tank with additions of two sources of
crude oils with and without dispersants, (3) a moored deployment in Baltimore Harbor, and (4)
hydrocast surveys in the Gulf of Mexico near a leaking oil barge.
In this Verification Statement, we present the performance results of the Chelsea
Technologies Group (CTG) UviLux Hydrocarbon (UV-HC) and CDOM (UV-CDOM) fluorometers.
Quality assurance (QA) oversight of the verification was provided by an ACT QA s.....
Resource URL
http://www.act-us.info/evaluations.phpPublisher
Alliance for Coastal Technologies (ACT)Solomons, MD
Series;Nr
ACT VS; 12-02Document Language
enEssential Ocean Variables (EOV)
Particulate matterBest Practice Type
Best PracticeStandard Operating Procedure
Citation
Johengen, T.; Smith, G.J.; Purcell, H.; Loranger, S.; Gilbert, S.; Maurer, T.; Gundersen, K.; Robertson, C.; Tamburri, M. (2013) Performance Verification Statement for the Chelsea UviLux Hydrocarbon and CDOM Fluorometers. Solomons, MD, Alliance for Coastal Technologies, 58pp. (ACTVS12-02). DOI: http://dx.doi.org/10.25607/OBP-332Collections
The following license files are associated with this item: