Performance Verification Statement for the Turner C3 Fluorometer

View/ Open
Average rating
votes
Date
2013Author
Johengen, T.
Smith, G.J.
Purcell, H.
Loranger, S.
Gilbert, S.
Maurer, T.
Gundersen, K.
Robertson, C.
Tamburri, M.
Status
PublishedPages
73pp.
Metadata
Show full item recordAbstract
Instrument performance verification is necessary so that effective existing technologies can
be recognized, and so that promising new technologies can become available to support coastal
science, resource management, and ocean observing systems. The Alliance for Coastal Technologies
(ACT) has therefore completed an evaluation of commercially available in situ hydrocarbon sensors.
This verification included test applications for: (1) controlled laboratory tanks with additions of
various organic, fluorescent compounds, (2) experimental wave tank with additions of two sources of
crude oils with and without dispersants, (3) a moored deployment in Baltimore Harbor, and (4)
hydrocast surveys in the Gulf of Mexico near a leaking oil barge.
In this Verification Statement, we present the performance results of a Turner Designs C3 in
situ fluorometer configured for concurrent measurement of refined fuel, crude oil, and CDOM levels.
The C3 package tested in this verification integrated .....
Resource URL
http://www.act-us.info/evaluations.phpPublisher
Alliance for Coastal Technologies (ACT)Solomons, MD
Series;Nr
ACT VS; 12-03Document Language
enEssential Ocean Variables (EOV)
Ocean ColourBest Practice Type
Best PracticeStandard Operating Procedure
Citation
Johengen, T.; Smith, G.J.; Purcell, H.; Loranger, S.; Gilbert, S.; Maurer, T.; Gundersen, K.; Robertson, C. and Tamburri, M. (2013) Performance Verification Statement for the Turner C3 Fluorometer. Solomons MD, Alliance for Coastal Technologies, 73pp. (ACTVS12-03). DOI: http://dx.doi.org/10.25607/OBP-321Collections