Performance Verification Statement for the In-Situ Troll 9500 pH Sensor.

View/ Open
Average rating
votes
Date
2015Author
Johengen, T.
Smith, G.J.
Schar, D.
Atkinson, M.
Purcell, H.
Loewensteiner, D.
Epperson, Z.
Tamburri, M.
Status
PublishedPages
58pp.
Metadata
Show full item recordAbstract
The Alliance for Coastal Technology (ACT) conducted a sensor verification study of in
situ pH sensors during 2013 and 2014 to characterize performance measures of accuracy and
reliability in a series of controlled laboratory studies and field mooring tests in diverse coastal
environments. A ten week long laboratory study was conducted at the Hawaii Institute of
Marine Biology and involved week long exposures at a full range of temperature and salinity
conditions. Tests were conducted at three fixed salinity levels (0.03, 22, 35) at each of three
fixed temperatures (10, 20, 30 oC). Ambient pH in the test tank was allowed to vary naturally
over the first five days. On the sixth day the pH was rapidly modified using acid/base additions
to compare accuracy over an extended range and during rapid changes. On the seventh day the
temperature was rapidly shifted to the next test condition. On the tenth week a repeated seawater
trial was conducted for two days while the temperature wa.....
Resource URL
http://www.act-us.info/evaluations.phpPublisher
Alliance for Coastal Technologies (ACT)Solomons, MD
Series;Nr
ACT VS; 15-04Document Language
enEssential Ocean Variables (EOV)
Inorganic carbonBest Practice Type
Best PracticeStandard Operating Procedure
Citation
Johengen, T.; Smith, G.J.; Schar, D.; Atkinson, M.; Purcell, H.; Loewensteiner, D.; Epperson, Z. and Tamburri, M. (2015) Performance Verification Statement for the In-Situ Troll 9500 pH Sensor. Solomons, MD, Alliance for Coastal Technologies, 58pp. (ACTVS15-04). http://dx.doi.org/10.25607/OBP-304Collections